Calculation of Depth-Dose Distribution of Intermediate Energy Heavy-Ion Beam
- Received Date: 1999-04-06
- Accepted Date: 1900-01-01
- Available Online: 2000-04-05
Abstract: A complex straggling algorithm is presented to calculate depth dose distribution of intermediate energy heavy-ion beam. Relative dose Bragg curves of carbon and oxygen ion beams with intermediate energies were computed according to the algorithm above. Comparisons between the relative doses predicted and measured were taken up. The Bragg curves predicted by the complex straggling algorithm show a good agreement with the experimental data. So the reasonableness of this algorithm and its reliability for calculating depth doses were verified. Hence the principle and method to deduce the depth dose distribution of therapeutic beam was integrated after this algorithm was developed.





Abstract
HTML
Reference
Related
PDF












DownLoad: