Diagnosis of the Output Characteristics of BSRF-3B3 Front End and Beamline
- Received Date: 2005-03-14
- Accepted Date: 1900-01-01
- Available Online: 2005-12-05
Abstract: All of the output characteristics of BSRF-3B3 beamline have achieved or surpassed the designed values after re-aligning and re-adjusting on-line. The flux is 7×1010phs/s/100mA and the best result of energy resolution (E/ΛE)is 5000@3.206keV at the station. The adjusting process, the diagnostic methods and the measured results of the output characteristics of both the front end and beamline are introduced in detail. The functions and shortages of double-wire BPM and fluorescence BPM for diagnosing and adjusting are analyzed.