A region segmentation based algorithm for building a crystal position lookup table in a scintillation detector

Get Citation
WANG Hai-Peng, YUN Ming-Kai, LIU Shuang-Quan, FAN Xin, CAO Xue-Xiang, CHAI Pei and SHAN Bao-Ci. A region segmentation based algorithm for building a crystal position lookup table in a scintillation detector[J]. Chinese Physics C, 2015, 39(3): 038201. doi: 10.1088/1674-1137/39/3/038201
WANG Hai-Peng, YUN Ming-Kai, LIU Shuang-Quan, FAN Xin, CAO Xue-Xiang, CHAI Pei and SHAN Bao-Ci. A region segmentation based algorithm for building a crystal position lookup table in a scintillation detector[J]. Chinese Physics C, 2015, 39(3): 038201.  doi: 10.1088/1674-1137/39/3/038201 shu
Milestone
Received: 2014-05-04
Revised: 2014-06-12
Article Metric

Article Views(1815)
PDF Downloads(211)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

A region segmentation based algorithm for building a crystal position lookup table in a scintillation detector

    Corresponding author: SHAN Bao-Ci,

Abstract: In a scintillation detector, scintillation crystals are typically made into a 2-dimensional modular array. The location of incident gamma-ray needs be calibrated due to spatial response nonlinearity. Generally, position histograms——the characteristic flood response of scintillation detectors-are used for position calibration. In this paper, a position calibration method based on a crystal position lookup table which maps the inaccurate location calculated by Anger logic to the exact hitting crystal position has been proposed. Firstly, the position histogram is preprocessed, such as noise reduction and image enhancement. Then the processed position histogram is segmented into disconnected regions, and crystal marking points are labeled by finding the centroids of regions. Finally, crystal boundaries are determined and the crystal position lookup table is generated. The scheme is evaluated by the whole-body positron emission tomography (PET) scanner and breast dedicated single photon emission computed tomography scanner developed by the Institute of High Energy Physics, Chinese Academy of Sciences. The results demonstrate that the algorithm is accurate, efficient, robust and applicable to any configurations of scintillation detector.