Design and experimental study of a two-dimensional position sensitive X-ray detector

Get Citation
WANG Xiao-Hu, ZHU Qi-Ming, CHEN Yuan-Bo, CHEN Chang, LIU Rong-Guang, WANG Lan, JIN Yan, MA Xiao-Yan and JIANG Shan. Design and experimental study of a two-dimensional position sensitive X-ray detector[J]. Chinese Physics C, 2008, 32(11): 903-907. doi: 10.1088/1674-1137/32/11/011
WANG Xiao-Hu, ZHU Qi-Ming, CHEN Yuan-Bo, CHEN Chang, LIU Rong-Guang, WANG Lan, JIN Yan, MA Xiao-Yan and JIANG Shan. Design and experimental study of a two-dimensional position sensitive X-ray detector[J]. Chinese Physics C, 2008, 32(11): 903-907.  doi: 10.1088/1674-1137/32/11/011 shu
Milestone
Received: 2008-02-25
Revised: 2008-03-17
Article Metric

Article Views(2807)
PDF Downloads(739)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Design and experimental study of a two-dimensional position sensitive X-ray detector

    Corresponding author: WANG Xiao-Hu,

Abstract: 

A prototype of a two-dimensional position sensitive X-ray detector was designed and constructed for small angle X-ray scattering experiments at BSFR (Beijing Synchrotron Radiation Facility). The detector is based on MWPC with cathode strip readout, and has a sensitive area of 200 mm×200 mm. The spatial resolution (FWHM) of about 210 μm along the anode wire direction was obtained from the 55Fe X-ray test of the detector.