Energy Calibration for a Plastic Scintillator Beta-Detector Telescope

Get Citation
Xu Shuwei and Liu Manqing. Energy Calibration for a Plastic Scintillator Beta-Detector Telescope[J]. Chinese Physics C, 1990, 14(S1): 101-104.
Xu Shuwei and Liu Manqing. Energy Calibration for a Plastic Scintillator Beta-Detector Telescope[J]. Chinese Physics C, 1990, 14(S1): 101-104. shu
Milestone
Received: 1989-02-27
Fund
Article Metric

Article Views(397)
PDF Downloads(9)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Energy Calibration for a Plastic Scintillator Beta-Detector Telescope

  • Institute of Modem Physics, Chinese Academy of Sciences, Lanzhou

Abstract: A plastic scintillator beta-detector telescope has been built and calibrated. Tests show that the uncertainty of linear energy calibration in the range of endpoint energy between 0.7 and 6.1 MeV for both β- and β+-emitters is 35 keV.