Design and preliminary test of a free-air ionization chamber for low-energy X-ray

Get Citation
WU Jin-Jie, YANG Yuan-Di, WANG Pei-Wei, CHEN Jing and LIU Jia-Cheng. Design and preliminary test of a free-air ionization chamber for low-energy X-ray[J]. Chinese Physics C, 2011, 35(6): 576-579. doi: 10.1088/1674-1137/35/6/013
WU Jin-Jie, YANG Yuan-Di, WANG Pei-Wei, CHEN Jing and LIU Jia-Cheng. Design and preliminary test of a free-air ionization chamber for low-energy X-ray[J]. Chinese Physics C, 2011, 35(6): 576-579.  doi: 10.1088/1674-1137/35/6/013 shu
Milestone
Received: 2010-08-31
Revised: 2010-10-11
Article Metric

Article Views(2414)
PDF Downloads(521)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Design and preliminary test of a free-air ionization chamber for low-energy X-ray

Abstract: A free-air ionization chamber in low-energy X-ray has been designed and manufactured at the National Institute of Metrology (NIM, China) according to the defination of air-kerma. The results of a preliminary test show that the leakage current of ionization chamber is around 2 × 10-15A, and the correction factor of ion recombination for the ionization chamber is also obtained. The free-air ionization chamber is suitable for the primary standard in low-energy X-rays.